Patient
- 72-year-old man
- Ingenio pacemaker (Boston Scientific)
- ATR episodes recorded in the device memory
Tracing
- oversensing of a very fast signal in the atrial channel (viewable at the ventricular channel level but not sensed) saturating the baseline
- entry counter filled (programmed at 8); duration begins; acceleration of the ventricular pacing rate
- switch to an asynchronous mode
Comments
- this tracing corresponds to the recording of a 50 Hz source and not to a real atrial arrhythmia
- it is essential to systematically verify all episodes recorded in the device memory
- analysis of the EGMs can sometimes correct an erroneous diagnosis concluded by the device
- the solution in this patient is to identify and eliminate the source of the interference